MajwareMAJWARE
Display

(0028,1053)

Rescale Slope

Keyword: RescaleSlope

The value m in the linear transformation: Output = m × SV + b. The slope of the linear mapping from stored values to output values.

Tag Attributes

Tag ID
(0028,1053)
Compact
00281053
Keyword
RescaleSlope
VR
DS — Decimal String
VM
1
Retired
No

Module(s)

Modality LUT Module

Engineering Notes

For standard CT, Rescale Slope = 1 and Intercept = -1024. A slope ≠ 1 is used for dose maps, PET SUV images, and some radiography workflows.

Troubleshooting

PET images showing values far out of expected SUV range may have wrong Rescale Slope. Verify against the scanner's quantification documentation.

Common Values

ValueMeaning
1Standard CT (1:1 mapping)
0.001PET SUV images (sub-unit scaling)