(0028,1050)

Window Center

Keyword: WindowCenter

Window Center for display: the centre of the range of values that is mapped to the full grayscale range of the display.

Tag Attributes

Tag ID
(0028,1050)
Compact
00281050
Keyword
WindowCenter
VR
DS — Decimal String
VM
1-n
Required
Type 1C — required when a condition applies
PS3.15
Not listed in PS3.15 Table E.1-1
Retired
No

Defined in

VOI LUT Module

What Window Center contains

Window Center and Window Width (0028,1051) choose which range of values fills the grayscale of the display. With the default LINEAR function, values below roughly center − width/2 show as black, values above roughly center + width/2 as white, and values in between as a ramp of grays.

The window applies to values after the Modality LUT, so for CT it is in Hounsfield units. A file can carry several center/width pairs, for example a soft-tissue and a bone window; Window Center & Width Explanation (0028,1055) names them, and VOI LUT Function (0028,1056) can select LINEAR_EXACT or SIGMOID instead of the default.

Presets such as brain (40/80), soft tissue (40/400) and lung (−600/1500) are clinical conventions, not values defined by DICOM, and they vary between sites.

It is applied after the Modality LUT (Rescale Slope and Intercept), so for CT it is in Hounsfield units. Several centre/width pairs can be stored, one per value, and Window Center & Width Explanation (0028,1055) can name each pair.

Group 0028 describes the pixel matrix and how stored pixel values are turned into displayed values.

Value format: DS (Decimal String), VM 1-n

(0028,1050) WindowCenter has Value Representation DS (Decimal String) and Value Multiplicity 1-n, which means one or more values, separated by a backslash (\).

DS — Decimal String

A fixed-point or floating-point number written as text, for example 0.5, -1024 or 1.2E-3. Leading and trailing spaces are allowed.

Length
16 bytes maximum
Characters
The digits 0–9, +, -, E, e, the period and the space.

Example value

40

Format mistakes to avoid

  • Writing a number at full floating-point precision can exceed 16 characters, which is invalid.
  • A comma as decimal separator (0,5) is invalid; DICOM always uses the period.

Common values

ValueMeaning
40With width 80: brain (CT, a common local convention, not a DICOM value)
40With width 400: soft tissue / abdomen (CT convention)
-600With width 1500: lung (CT convention)
400With width 1800: bone (CT convention)

Where WindowCenter appears

Defined inTypeMeaning
VOI LUT ModulePS3.3 C.11.21CRequired, with a value, when its condition applies, and absent otherwise.Required if VOI LUT Sequence (0028,3010) is not present; may be present otherwise.

A module lists the attributes an object must or may carry. Many IODs (CT Image, MR Image, Secondary Capture and others) include the same module, and an IOD-specific module can make an attribute stricter than the general one.

De-identification (PS3.15)

Window Center is not listed in PS3.15 Table E.1-1, so the Basic Application Level Confidentiality Profile gives no specific action for it.

To see how de-identification changes your own files, try the free DICOM Anonymizer, which runs in the browser and uploads nothing.

In archives and migrations

Display attributes are copied unchanged in a migration. Viewers may still render the same images differently, because each viewer decides how it applies stored windows and lookup tables.

Common errors and how to fix them

  • The image opens completely black or white

    The window does not cover the data, for example a window in HU applied to values before the rescale. Check the stored window against the rescaled value range.

  • Window Center and Window Width have different numbers of values

    They form pairs, so both need the same number of values. Fix the writer.

Troubleshooting: If images open too dark or too bright, check the stored window, and whether the viewer uses it or its own preset. Window Width must accompany Window Center, with the same number of values.

Check a file in seconds with the free DICOM Validator: it reads the header in your browser, lists the tags with their values, and flags missing or malformed key attributes.

Window Center: frequently asked questions

What is DICOM tag (0028,1050) Window Center?

Window Center for display: the centre of the range of values that is mapped to the full grayscale range of the display. Window Center and Window Width (0028,1051) choose which range of values fills the grayscale of the display. With the default LINEAR function, values below roughly center − width/2 show as black, values above roughly center + width/2 as white, and values in between as a ramp of grays.

What VR and VM does Window Center (0028,1050) have?

(0028,1050) WindowCenter has Value Representation DS (Decimal String) and Value Multiplicity 1-n, which means one or more values, separated by a backslash (\). A fixed-point or floating-point number written as text, for example 0.5, -1024 or 1.2E-3. Leading and trailing spaces are allowed. Maximum length: 16 bytes maximum.

Is Window Center (0028,1050) required?

In the VOI LUT Module (PS3.3 C.11.2) it is Type 1C: required, with a value, when its condition applies, and absent otherwise. Required if VOI LUT Sequence (0028,3010) is not present; may be present otherwise.

How is Window Center handled when de-identifying DICOM files?

Window Center is not listed in PS3.15 Table E.1-1, so the Basic Application Level Confidentiality Profile gives no specific action for it.

How do I troubleshoot Window Center (0028,1050)?

If images open too dark or too bright, check the stored window, and whether the viewer uses it or its own preset. Window Width must accompany Window Center, with the same number of values.

What window settings are used for CT lung?

A common lung window has a center of about −600 HU and a width of about 1500 HU. These are local clinical conventions, not DICOM values; Window Center (0028,1050) and Window Width (0028,1051) in the image are only the defaults the modality stored.

Check WindowCenter in your files

The free DICOM Validator lists every header tag and flags problems. Files stay in your browser.

Open DICOM Validator →

Remove patient data before sharing

The free DICOM Anonymizer applies a de-identification profile in your browser, with no upload.