MajwareMAJWARE
Display

(0028,1052)

Rescale Intercept

Keyword: RescaleIntercept

The value b in the linear transformation from stored pixel values to output units: Output = m × SV + b.

Tag Attributes

Tag ID
(0028,1052)
Compact
00281052
Keyword
RescaleIntercept
VR
DS — Decimal String
VM
1
Retired
No

Module(s)

Modality LUT Module

Engineering Notes

For CT, Rescale Intercept is −1024 to convert stored integers to Hounsfield Units. Incorrect value shifts all HU values, causing systematic measurement errors.

Troubleshooting

CT Hounsfield Unit values systematically off by a constant (e.g., water reads as +24 instead of 0) — check Rescale Intercept and Slope. Should be -1024 and 1 respectively for standard CT.

Common Values

ValueMeaning
-1024CT — converts to Hounsfield Units
0No offset (most non-CT modalities)