MajwareMAJWARE
Series

(0020,0060)

Laterality

Keyword: Laterality

Laterality of the paired body part examined in the series. Enumerated Values: R (right) and L (left).

Tag Attributes

Tag ID
(0020,0060)
Compact
00200060
Keyword
Laterality
VR
CS — Code String
VM
1
Required
Type 2C — required when a condition applies, may be empty
PS3.15
Not listed in PS3.15 Table E.1-1
Retired
No

Defined in

General Series Module

What Laterality contains

There is no value for both sides: when a series covers both, Laterality is not used at series level and the side is given per image in Image Laterality (0020,0062). A wrong side is a patient-safety problem.

Group 0020 holds the identifiers and numbers that tie instances into series and studies, and the geometry of the patient coordinate system.

Value format: CS (Code String), VM 1

(0020,0060) Laterality has Value Representation CS (Code String) and Value Multiplicity 1, which means exactly one value.

CS — Code String

A short code. Leading and trailing spaces are not significant. Codes are usually Defined Terms or Enumerated Values listed by the standard.

Length
16 bytes maximum
Characters
Uppercase letters A–Z, digits 0–9, the space and the underscore.

Example value

R

Format mistakes to avoid

  • Lowercase letters, hyphens and other punctuation are not allowed, so ct or Head-Neck are invalid code strings.
  • Values longer than 16 characters are invalid.

Common values

ValueMeaning
RRight
LLeft

Where Laterality appears

Defined inTypeMeaning
General Series ModulePS3.3 C.7.3.12CRequired, possibly empty, when its condition applies.Required if the body part examined is paired and Image Laterality (0020,0062), Frame Laterality (0020,9072) or Measurement Laterality (0024,0113) is not sent.

A module lists the attributes an object must or may carry. Many IODs (CT Image, MR Image, Secondary Capture and others) include the same module, and an IOD-specific module can make an attribute stricter than the general one.

De-identification (PS3.15)

Laterality is not listed in PS3.15 Table E.1-1, so the Basic Application Level Confidentiality Profile gives no specific action for it.

To see how de-identification changes your own files, try the free DICOM Anonymizer, which runs in the browser and uploads nothing.

In archives and migrations

Series-level attributes group and order images in viewers. Copy them unchanged in a migration, and compare series counts and series descriptions between the old and the new archive when checking that a study arrived complete.

Common errors and how to fix them

Troubleshooting: Values such as B, U or lowercase letters are invalid here (B and U belong to Image Laterality). If images of both sides share a series, send Image Laterality per image and leave Laterality out.

Check a file in seconds with the free DICOM Validator: it reads the header in your browser, lists the tags with their values, and flags missing or malformed key attributes.

Laterality: frequently asked questions

What is DICOM tag (0020,0060) Laterality?

Laterality of the paired body part examined in the series. Enumerated Values: R (right) and L (left).

What VR and VM does Laterality (0020,0060) have?

(0020,0060) Laterality has Value Representation CS (Code String) and Value Multiplicity 1, which means exactly one value. A short code. Leading and trailing spaces are not significant. Codes are usually Defined Terms or Enumerated Values listed by the standard. Maximum length: 16 bytes maximum.

Is Laterality (0020,0060) required?

In the General Series Module (PS3.3 C.7.3.1) it is Type 2C: required, possibly empty, when its condition applies. Required if the body part examined is paired and Image Laterality (0020,0062), Frame Laterality (0020,9072) or Measurement Laterality (0024,0113) is not sent.

How is Laterality handled when de-identifying DICOM files?

Laterality is not listed in PS3.15 Table E.1-1, so the Basic Application Level Confidentiality Profile gives no specific action for it.

How do I troubleshoot Laterality (0020,0060)?

Values such as B, U or lowercase letters are invalid here (B and U belong to Image Laterality). If images of both sides share a series, send Image Laterality per image and leave Laterality out.

Check Laterality in your files

The free DICOM Validator lists every header tag and flags problems. Files stay in your browser.

Open DICOM Validator →

Remove patient data before sharing

The free DICOM Anonymizer applies a de-identification profile in your browser, with no upload.

Open DICOM Anonymizer →

Sources

Checked against the DICOM standard, release 2026d. Module Types, query keys and de-identification actions come from these tables:

More Series tags