(0008,0060)

Modality

Keyword: Modality

Type of equipment that originally acquired the data. Defines the imaging modality of the Series.

Tag Attributes

Tag ID
(0008,0060)
Compact
00080060
Keyword
Modality
VR
CS — Code String
VM
1
Required
Type 1 — required, with a value
PS3.15
Not listed in PS3.15 Table E.1-1
Retired
No

Defined in

General Series ModuleSC Equipment Module

What Modality contains

Modality is a code string, set per series, that names the type of equipment or object: CT, MR, US, CR, DX and others for images, and codes such as SR, PR, KO, SEG and DOC for non-image objects. It is Type 1 in the General Series Module, so every series must have one.

Because it is a series-level attribute, one study can contain several modalities, for example the CT images, a PR series of annotations and an SR dose report. Modalities in Study (0008,0061) lists them at study level.

Modality is a Required Key at series level in C-FIND and a Required matching key in Modality Worklist queries, where a device usually asks only for orders of its own modality.

Display protocols, hanging protocols, routing rules and worklist filters all key on it. It is set per series, so one study can hold CT, PR, SR and KO series side by side.

Group 0008 holds general identifying information: what kind of object this is, when it was made, and by which equipment and people.

Value format: CS (Code String), VM 1

(0008,0060) Modality has Value Representation CS (Code String) and Value Multiplicity 1, which means exactly one value.

CS — Code String

A short code. Leading and trailing spaces are not significant. Codes are usually Defined Terms or Enumerated Values listed by the standard.

Length
16 bytes maximum
Characters
Uppercase letters A–Z, digits 0–9, the space and the underscore.

Example value

CT

Format mistakes to avoid

  • Lowercase letters, hyphens and other punctuation are not allowed, so ct or Head-Neck are invalid code strings.
  • Values longer than 16 characters are invalid.

Common values

ValueMeaning
CRComputed Radiography
CTComputed Tomography
MRMagnetic Resonance
USUltrasound
NMNuclear Medicine
PTPositron Emission Tomography
DXDigital Radiography
MGMammography
RFRadio Fluoroscopy
XAX-Ray Angiography
OTOther
SCSecondary Capture
PRPresentation State
SRStructured Report
KOKey Object Selection
RTIMAGERadiotherapy Image
RTSTRUCTRadiotherapy Structure Set
SEGSegmentation
DOCDocument

Where Modality appears

Defined inTypeMeaning
General Series ModulePS3.3 C.7.3.11Required, and it must have a value.
SC Equipment ModulePS3.3 C.8.6.13Optional.In Secondary Capture images it is optional and overrides the General Series Module.

A module lists the attributes an object must or may carry. Many IODs (CT Image, MR Image, Secondary Capture and others) include the same module, and an IOD-specific module can make an attribute stricter than the general one.

In a C-FIND query using the Study Root Query/Retrieve Information Model, it is a SERIES-level key of type R: a Required Key: every Query/Retrieve SCP must support matching on it.

In a Modality Worklist query it is a matching key of type R (the worklist SCP must support matching on it) and a return key of type 1 (the SCP must return it with a value).

De-identification (PS3.15)

Modality is not listed in PS3.15 Table E.1-1, so the Basic Application Level Confidentiality Profile gives no specific action for it.

To see how de-identification changes your own files, try the free DICOM Anonymizer, which runs in the browser and uploads nothing.

In archives and migrations

Series-level attributes group and order images in viewers. Copy them unchanged in a migration, and compare series counts and series descriptions between the old and the new archive when checking that a study arrived complete.

Common errors and how to fix them

  • The modality worklist is empty or shows the wrong orders

    The Modality value in the worklist query does not match the value the RIS stored for the orders. Compare the C-FIND request with the RIS configuration.

  • The wrong hanging protocol or routing rule is applied

    The series carries an unexpected Modality, for example OT for images a rule expects as CT. Fix the value at the device that creates the series.

Troubleshooting: A wrong Modality applies the wrong hanging protocol or routing rule, for example CT presets on MR images. Check the value at the device that creates the series. For a study-level summary, query Modalities in Study (0008,0061).

Check a file in seconds with the free DICOM Validator: it reads the header in your browser, lists the tags with their values, and flags missing or malformed key attributes.

Modality: frequently asked questions

What is DICOM tag (0008,0060) Modality?

Type of equipment that originally acquired the data. Defines the imaging modality of the Series. Modality is a code string, set per series, that names the type of equipment or object: CT, MR, US, CR, DX and others for images, and codes such as SR, PR, KO, SEG and DOC for non-image objects. It is Type 1 in the General Series Module, so every series must have one.

What VR and VM does Modality (0008,0060) have?

(0008,0060) Modality has Value Representation CS (Code String) and Value Multiplicity 1, which means exactly one value. A short code. Leading and trailing spaces are not significant. Codes are usually Defined Terms or Enumerated Values listed by the standard. Maximum length: 16 bytes maximum.

Is Modality (0008,0060) required?

In the General Series Module (PS3.3 C.7.3.1) it is Type 1: required, and it must have a value. In the SC Equipment Module (PS3.3 C.8.6.1) it is Type 3: optional. In Secondary Capture images it is optional and overrides the General Series Module.

Can you query a PACS by Modality?

In a C-FIND query using the Study Root Query/Retrieve Information Model, it is a SERIES-level key of type R: a Required Key: every Query/Retrieve SCP must support matching on it. In a Modality Worklist query it is a matching key of type R (the worklist SCP must support matching on it) and a return key of type 1 (the SCP must return it with a value).

How is Modality handled when de-identifying DICOM files?

Modality is not listed in PS3.15 Table E.1-1, so the Basic Application Level Confidentiality Profile gives no specific action for it.

How do I troubleshoot Modality (0008,0060)?

A wrong Modality applies the wrong hanging protocol or routing rule, for example CT presets on MR images. Check the value at the device that creates the series. For a study-level summary, query Modalities in Study (0008,0061).

What is the DICOM modality code for PET?

PT. Other common codes are CT (computed tomography), MR (magnetic resonance), US (ultrasound), NM (nuclear medicine), CR and DX (radiography), MG (mammography) and XA (X-ray angiography).

Check Modality in your files

The free DICOM Validator lists every header tag and flags problems. Files stay in your browser.

Open DICOM Validator →

Remove patient data before sharing

The free DICOM Anonymizer applies a de-identification profile in your browser, with no upload.

Open DICOM Anonymizer →

Sources

More Series tags