Acquisition

(0018,1111)

Distance Source to Patient

Keyword: DistanceSourceToPatient

Distance in mm from the X-ray source to the isocentre (the centre of the field of view), traditionally called the Source Object Distance (SOD).

Tag Attributes

Tag ID
(0018,1111)
Compact
00181111
Keyword
DistanceSourceToPatient
VR
DS — Decimal String
VM
1
Required
Type 3 — optional
PS3.15
Not listed in PS3.15 Table E.1-1
Retired
No

Defined in

CT Image Module

What Distance Source to Patient contains

With Distance Source to Detector it gives the geometric magnification: SID ÷ SOD.

Group 0018 holds acquisition information: how the data were acquired and with which settings.

Value format: DS (Decimal String), VM 1

(0018,1111) DistanceSourceToPatient has Value Representation DS (Decimal String) and Value Multiplicity 1, which means exactly one value.

DS — Decimal String

A fixed-point or floating-point number written as text, for example 0.5, -1024 or 1.2E-3. Leading and trailing spaces are allowed.

Length
16 bytes maximum
Characters
The digits 0–9, +, -, E, e, the period and the space.

Example value

600

Format mistakes to avoid

  • Writing a number at full floating-point precision can exceed 16 characters, which is invalid.
  • A comma as decimal separator (0,5) is invalid; DICOM always uses the period.

Where DistanceSourceToPatient appears

Defined inTypeMeaning
CT Image ModulePS3.3 C.8.2.13Optional.

A module lists the attributes an object must or may carry. Many IODs (CT Image, MR Image, Secondary Capture and others) include the same module, and an IOD-specific module can make an attribute stricter than the general one.

De-identification (PS3.15)

Distance Source to Patient is not listed in PS3.15 Table E.1-1, so the Basic Application Level Confidentiality Profile gives no specific action for it.

To see how de-identification changes your own files, try the free DICOM Anonymizer, which runs in the browser and uploads nothing.

In archives and migrations

Acquisition parameters describe what happened at the scanner and should never change after the fact. A migration copies them as they are; a difference between the source and the copy points to a fault in the transfer, not to a correction.

Common errors and how to fix them

Troubleshooting: It is measured to the isocentre, not to the patient's skin, so it cannot be used directly as a skin distance in dose calculations.

Check a file in seconds with the free DICOM Validator: it reads the header in your browser, lists the tags with their values, and flags missing or malformed key attributes.

Distance Source to Patient: frequently asked questions

What is DICOM tag (0018,1111) Distance Source to Patient?

Distance in mm from the X-ray source to the isocentre (the centre of the field of view), traditionally called the Source Object Distance (SOD).

What VR and VM does Distance Source to Patient (0018,1111) have?

(0018,1111) DistanceSourceToPatient has Value Representation DS (Decimal String) and Value Multiplicity 1, which means exactly one value. A fixed-point or floating-point number written as text, for example 0.5, -1024 or 1.2E-3. Leading and trailing spaces are allowed. Maximum length: 16 bytes maximum.

Is Distance Source to Patient (0018,1111) required?

In the CT Image Module (PS3.3 C.8.2.1) it is Type 3: optional.

How is Distance Source to Patient handled when de-identifying DICOM files?

Distance Source to Patient is not listed in PS3.15 Table E.1-1, so the Basic Application Level Confidentiality Profile gives no specific action for it.

How do I troubleshoot Distance Source to Patient (0018,1111)?

It is measured to the isocentre, not to the patient's skin, so it cannot be used directly as a skin distance in dose calculations.

Check DistanceSourceToPatient in your files

The free DICOM Validator lists every header tag and flags problems. Files stay in your browser.

Open DICOM Validator →

Remove patient data before sharing

The free DICOM Anonymizer applies a de-identification profile in your browser, with no upload.

Open DICOM Anonymizer →

Sources

Checked against the DICOM standard, release 2026d. Module Types, query keys and de-identification actions come from these tables:

More Acquisition tags