(0018,1110)
Distance Source to Detector
Keyword: DistanceSourceToDetector
Distance in mm from the X-ray source to the centre of the detector, traditionally called the Source Image Receptor Distance (SID).
Tag Attributes
- Tag ID
- (0018,1110)
- Compact
- 00181110
- Keyword
- DistanceSourceToDetector
- VR
- DS — Decimal String
- VM
- 1
- Required
- Type 3 — optional
- PS3.15
- Not listed in PS3.15 Table E.1-1
- Retired
- No
Defined in
What Distance Source to Detector contains
Divided by Distance Source to Patient it gives the geometric magnification at the isocentre. The same attribute also appears in projection X-ray IODs.
Group 0018 holds acquisition information: how the data were acquired and with which settings.
Value format: DS (Decimal String), VM 1
(0018,1110) DistanceSourceToDetector has Value Representation DS (Decimal String) and Value Multiplicity 1, which means exactly one value.
DS — Decimal String
A fixed-point or floating-point number written as text, for example 0.5, -1024 or 1.2E-3. Leading and trailing spaces are allowed.
- Length
- 16 bytes maximum
- Characters
- The digits 0–9, +, -, E, e, the period and the space.
Example value
1100
Format mistakes to avoid
- Writing a number at full floating-point precision can exceed 16 characters, which is invalid.
- A comma as decimal separator (0,5) is invalid; DICOM always uses the period.
Where DistanceSourceToDetector appears
| Defined in | Type | Meaning |
|---|---|---|
| CT Image ModulePS3.3 C.8.2.1 | 3 | Optional. |
A module lists the attributes an object must or may carry. Many IODs (CT Image, MR Image, Secondary Capture and others) include the same module, and an IOD-specific module can make an attribute stricter than the general one.
De-identification (PS3.15)
Distance Source to Detector is not listed in PS3.15 Table E.1-1, so the Basic Application Level Confidentiality Profile gives no specific action for it.
To see how de-identification changes your own files, try the free DICOM Anonymizer, which runs in the browser and uploads nothing.
In archives and migrations
Acquisition parameters describe what happened at the scanner and should never change after the fact. A migration copies them as they are; a difference between the source and the copy points to a fault in the transfer, not to a correction.
Common errors and how to fix them
Troubleshooting: In projection radiography, measurements in the patient plane depend on magnification. If calipers read wrong, first check which pixel spacing the viewer uses (Pixel Spacing (0028,0030) or Imager Pixel Spacing (0018,1164)).
Check a file in seconds with the free DICOM Validator: it reads the header in your browser, lists the tags with their values, and flags missing or malformed key attributes.
Distance Source to Detector: frequently asked questions
What is DICOM tag (0018,1110) Distance Source to Detector?
Distance in mm from the X-ray source to the centre of the detector, traditionally called the Source Image Receptor Distance (SID).
What VR and VM does Distance Source to Detector (0018,1110) have?
(0018,1110) DistanceSourceToDetector has Value Representation DS (Decimal String) and Value Multiplicity 1, which means exactly one value. A fixed-point or floating-point number written as text, for example 0.5, -1024 or 1.2E-3. Leading and trailing spaces are allowed. Maximum length: 16 bytes maximum.
Is Distance Source to Detector (0018,1110) required?
In the CT Image Module (PS3.3 C.8.2.1) it is Type 3: optional.
How is Distance Source to Detector handled when de-identifying DICOM files?
Distance Source to Detector is not listed in PS3.15 Table E.1-1, so the Basic Application Level Confidentiality Profile gives no specific action for it.
How do I troubleshoot Distance Source to Detector (0018,1110)?
In projection radiography, measurements in the patient plane depend on magnification. If calipers read wrong, first check which pixel spacing the viewer uses (Pixel Spacing (0028,0030) or Imager Pixel Spacing (0018,1164)).
Check DistanceSourceToDetector in your files
The free DICOM Validator lists every header tag and flags problems. Files stay in your browser.
Open DICOM Validator →Remove patient data before sharing
The free DICOM Anonymizer applies a de-identification profile in your browser, with no upload.
Open DICOM Anonymizer →Sources
Checked against the DICOM standard, release 2026d. Module Types, query keys and de-identification actions come from these tables: