MajwareMAJWARE
Acquisition

(0018,1314)

Flip Angle

Keyword: FlipAngle

The flip angle in degrees for the MR pulse sequence.

Tag Attributes

Tag ID
(0018,1314)
Compact
00181314
Keyword
FlipAngle
VR
DS — Decimal String
VM
1
Required
Type 3 — optional
PS3.15
Not listed in PS3.15 Table E.1-1
Retired
No

Defined in

MR Image Module

What Flip Angle contains

Fundamental to MR sequence classification. Together with TR and TE, identifies the contrast mechanism (gradient echo, SSFP, etc.).

Group 0018 holds acquisition information: how the data were acquired and with which settings.

Value format: DS (Decimal String), VM 1

(0018,1314) FlipAngle has Value Representation DS (Decimal String) and Value Multiplicity 1, which means exactly one value.

DS — Decimal String

A fixed-point or floating-point number written as text, for example 0.5, -1024 or 1.2E-3. Leading and trailing spaces are allowed.

Length
16 bytes maximum
Characters
The digits 0–9, +, -, E, e, the period and the space.

Example value

90

Format mistakes to avoid

  • Writing a number at full floating-point precision can exceed 16 characters, which is invalid.
  • A comma as decimal separator (0,5) is invalid; DICOM always uses the period.

Where FlipAngle appears

Defined inTypeMeaning
MR Image ModulePS3.3 C.8.3.13Optional.

A module lists the attributes an object must or may carry. Many IODs (CT Image, MR Image, Secondary Capture and others) include the same module, and an IOD-specific module can make an attribute stricter than the general one.

De-identification (PS3.15)

Flip Angle is not listed in PS3.15 Table E.1-1, so the Basic Application Level Confidentiality Profile gives no specific action for it.

To see how de-identification changes your own files, try the free DICOM Anonymizer, which runs in the browser and uploads nothing.

In archives and migrations

Acquisition parameters describe what happened at the scanner and should never change after the fact. A migration copies them as they are; a difference between the source and the copy points to a fault in the transfer, not to a correction.

Common errors and how to fix them

Troubleshooting: Optional (Type 3). Software that classifies sequences should not require it, and should combine it with Scanning Sequence and the timing parameters.

Check a file in seconds with the free DICOM Validator: it reads the header in your browser, lists the tags with their values, and flags missing or malformed key attributes.

Flip Angle: frequently asked questions

What is DICOM tag (0018,1314) Flip Angle?

The flip angle in degrees for the MR pulse sequence.

What VR and VM does Flip Angle (0018,1314) have?

(0018,1314) FlipAngle has Value Representation DS (Decimal String) and Value Multiplicity 1, which means exactly one value. A fixed-point or floating-point number written as text, for example 0.5, -1024 or 1.2E-3. Leading and trailing spaces are allowed. Maximum length: 16 bytes maximum.

Is Flip Angle (0018,1314) required?

In the MR Image Module (PS3.3 C.8.3.1) it is Type 3: optional.

How is Flip Angle handled when de-identifying DICOM files?

Flip Angle is not listed in PS3.15 Table E.1-1, so the Basic Application Level Confidentiality Profile gives no specific action for it.

How do I troubleshoot Flip Angle (0018,1314)?

Optional (Type 3). Software that classifies sequences should not require it, and should combine it with Scanning Sequence and the timing parameters.

Check FlipAngle in your files

The free DICOM Validator lists every header tag and flags problems. Files stay in your browser.

Open DICOM Validator →

Remove patient data before sharing

The free DICOM Anonymizer applies a de-identification profile in your browser, with no upload.

Open DICOM Anonymizer →

Sources

Checked against the DICOM standard, release 2026d. Module Types, query keys and de-identification actions come from these tables: