Acquisition

(0018,1151)

X-Ray Tube Current

Keyword: XRayTubeCurrent

X-ray tube current, in mA.

Tag Attributes

Tag ID
(0018,1151)
Compact
00181151
Keyword
XRayTubeCurrent
VR
IS — Integer String
VM
1
Required
Type 3 — optional
PS3.15
Not listed in PS3.15 Table E.1-1
Retired
No

Defined in

CT Image Module

What X-Ray Tube Current contains

With tube current modulation the current changes during the scan, so one value per image is at best a summary. Use the Radiation Dose Structured Report for dose figures.

Group 0018 holds acquisition information: how the data were acquired and with which settings.

Value format: IS (Integer String), VM 1

(0018,1151) XRayTubeCurrent has Value Representation IS (Integer String) and Value Multiplicity 1, which means exactly one value.

IS — Integer String

A whole number written as text, from −2³¹ to 2³¹−1. Leading and trailing spaces are allowed.

Length
12 bytes maximum
Characters
The digits 0–9, + and -, and the space.

Example value

250

Format mistakes to avoid

  • A decimal point is not allowed, so 12.0 is not a valid Integer String.

Where XRayTubeCurrent appears

Defined inTypeMeaning
CT Image ModulePS3.3 C.8.2.13Optional.

A module lists the attributes an object must or may carry. Many IODs (CT Image, MR Image, Secondary Capture and others) include the same module, and an IOD-specific module can make an attribute stricter than the general one.

De-identification (PS3.15)

X-Ray Tube Current is not listed in PS3.15 Table E.1-1, so the Basic Application Level Confidentiality Profile gives no specific action for it.

To see how de-identification changes your own files, try the free DICOM Anonymizer, which runs in the browser and uploads nothing.

In archives and migrations

Acquisition parameters describe what happened at the scanner and should never change after the fact. A migration copies them as they are; a difference between the source and the copy points to a fault in the transfer, not to a correction.

Common errors and how to fix them

Troubleshooting: Optional (Type 3) in CT images. Software that needs the tube current should handle its absence rather than treat it as an error.

Check a file in seconds with the free DICOM Validator: it reads the header in your browser, lists the tags with their values, and flags missing or malformed key attributes.

X-Ray Tube Current: frequently asked questions

What is DICOM tag (0018,1151) X-Ray Tube Current?

X-ray tube current, in mA.

What VR and VM does X-Ray Tube Current (0018,1151) have?

(0018,1151) XRayTubeCurrent has Value Representation IS (Integer String) and Value Multiplicity 1, which means exactly one value. A whole number written as text, from −2³¹ to 2³¹−1. Leading and trailing spaces are allowed. Maximum length: 12 bytes maximum.

Is X-Ray Tube Current (0018,1151) required?

In the CT Image Module (PS3.3 C.8.2.1) it is Type 3: optional.

How is X-Ray Tube Current handled when de-identifying DICOM files?

X-Ray Tube Current is not listed in PS3.15 Table E.1-1, so the Basic Application Level Confidentiality Profile gives no specific action for it.

How do I troubleshoot X-Ray Tube Current (0018,1151)?

Optional (Type 3) in CT images. Software that needs the tube current should handle its absence rather than treat it as an error.

Check XRayTubeCurrent in your files

The free DICOM Validator lists every header tag and flags problems. Files stay in your browser.

Open DICOM Validator →

Remove patient data before sharing

The free DICOM Anonymizer applies a de-identification profile in your browser, with no upload.

Open DICOM Anonymizer →

Sources

Checked against the DICOM standard, release 2026d. Module Types, query keys and de-identification actions come from these tables:

More Acquisition tags