Acquisition

(0018,0021)

Sequence Variant

Keyword: SequenceVariant

Variants of the scanning sequence, such as segmented k-space, spoiled or magnetization-prepared.

Tag Attributes

Tag ID
(0018,0021)
Compact
00180021
Keyword
SequenceVariant
VR
CS — Code String
VM
1-n
Required
Type 1 — required, with a value
PS3.15
Not listed in PS3.15 Table E.1-1
Retired
No

Defined in

MR Image Module

What Sequence Variant contains

Specifies modifications to the basic scanning sequence. Required for complete MR sequence identification.

Group 0018 holds acquisition information: how the data were acquired and with which settings.

Value format: CS (Code String), VM 1-n

(0018,0021) SequenceVariant has Value Representation CS (Code String) and Value Multiplicity 1-n, which means one or more values, separated by a backslash (\).

CS — Code String

A short code. Leading and trailing spaces are not significant. Codes are usually Defined Terms or Enumerated Values listed by the standard.

Length
16 bytes maximum
Characters
Uppercase letters A–Z, digits 0–9, the space and the underscore.

Example value

SK\SP

Format mistakes to avoid

  • Lowercase letters, hyphens and other punctuation are not allowed, so ct or Head-Neck are invalid code strings.
  • Values longer than 16 characters are invalid.

Common values

ValueMeaning
SKSegmented k-space
MTCMagnetisation Transfer Contrast
SSSteady State
TRSSTime Reversed Steady State
SPSpoiled
MPMAG Prepared
OSPOversampling Phase
NONENo sequence variant

Where SequenceVariant appears

Defined inTypeMeaning
MR Image ModulePS3.3 C.8.3.11Required, and it must have a value.

A module lists the attributes an object must or may carry. Many IODs (CT Image, MR Image, Secondary Capture and others) include the same module, and an IOD-specific module can make an attribute stricter than the general one.

De-identification (PS3.15)

Sequence Variant is not listed in PS3.15 Table E.1-1, so the Basic Application Level Confidentiality Profile gives no specific action for it.

To see how de-identification changes your own files, try the free DICOM Anonymizer, which runs in the browser and uploads nothing.

In archives and migrations

Acquisition parameters describe what happened at the scanner and should never change after the fact. A migration copies them as they are; a difference between the source and the copy points to a fault in the transfer, not to a correction.

Common errors and how to fix them

Troubleshooting: NONE is a valid value and means no variant applies; it is not the same as a missing attribute, which makes the MR image invalid because the attribute is Type 1.

Check a file in seconds with the free DICOM Validator: it reads the header in your browser, lists the tags with their values, and flags missing or malformed key attributes.

Sequence Variant: frequently asked questions

What is DICOM tag (0018,0021) Sequence Variant?

Variants of the scanning sequence, such as segmented k-space, spoiled or magnetization-prepared.

What VR and VM does Sequence Variant (0018,0021) have?

(0018,0021) SequenceVariant has Value Representation CS (Code String) and Value Multiplicity 1-n, which means one or more values, separated by a backslash (\). A short code. Leading and trailing spaces are not significant. Codes are usually Defined Terms or Enumerated Values listed by the standard. Maximum length: 16 bytes maximum.

Is Sequence Variant (0018,0021) required?

In the MR Image Module (PS3.3 C.8.3.1) it is Type 1: required, and it must have a value.

How is Sequence Variant handled when de-identifying DICOM files?

Sequence Variant is not listed in PS3.15 Table E.1-1, so the Basic Application Level Confidentiality Profile gives no specific action for it.

How do I troubleshoot Sequence Variant (0018,0021)?

NONE is a valid value and means no variant applies; it is not the same as a missing attribute, which makes the MR image invalid because the attribute is Type 1.

Check SequenceVariant in your files

The free DICOM Validator lists every header tag and flags problems. Files stay in your browser.

Open DICOM Validator →

Remove patient data before sharing

The free DICOM Anonymizer applies a de-identification profile in your browser, with no upload.

Open DICOM Anonymizer →

Sources

Checked against the DICOM standard, release 2026d. Module Types, query keys and de-identification actions come from these tables:

More Acquisition tags