(0018,0021)
Sequence Variant
Keyword: SequenceVariant
Variants of the scanning sequence, such as segmented k-space, spoiled or magnetization-prepared.
Tag Attributes
- Tag ID
- (0018,0021)
- Compact
- 00180021
- Keyword
- SequenceVariant
- VR
- CS — Code String
- VM
- 1-n
- Required
- Type 1 — required, with a value
- PS3.15
- Not listed in PS3.15 Table E.1-1
- Retired
- No
Defined in
What Sequence Variant contains
Specifies modifications to the basic scanning sequence. Required for complete MR sequence identification.
Group 0018 holds acquisition information: how the data were acquired and with which settings.
Value format: CS (Code String), VM 1-n
(0018,0021) SequenceVariant has Value Representation CS (Code String) and Value Multiplicity 1-n, which means one or more values, separated by a backslash (\).
CS — Code String
A short code. Leading and trailing spaces are not significant. Codes are usually Defined Terms or Enumerated Values listed by the standard.
- Length
- 16 bytes maximum
- Characters
- Uppercase letters A–Z, digits 0–9, the space and the underscore.
Example value
SK\SP
Format mistakes to avoid
- Lowercase letters, hyphens and other punctuation are not allowed, so ct or Head-Neck are invalid code strings.
- Values longer than 16 characters are invalid.
Common values
| Value | Meaning |
|---|---|
| SK | Segmented k-space |
| MTC | Magnetisation Transfer Contrast |
| SS | Steady State |
| TRSS | Time Reversed Steady State |
| SP | Spoiled |
| MP | MAG Prepared |
| OSP | Oversampling Phase |
| NONE | No sequence variant |
Where SequenceVariant appears
| Defined in | Type | Meaning |
|---|---|---|
| MR Image ModulePS3.3 C.8.3.1 | 1 | Required, and it must have a value. |
A module lists the attributes an object must or may carry. Many IODs (CT Image, MR Image, Secondary Capture and others) include the same module, and an IOD-specific module can make an attribute stricter than the general one.
De-identification (PS3.15)
Sequence Variant is not listed in PS3.15 Table E.1-1, so the Basic Application Level Confidentiality Profile gives no specific action for it.
To see how de-identification changes your own files, try the free DICOM Anonymizer, which runs in the browser and uploads nothing.
In archives and migrations
Acquisition parameters describe what happened at the scanner and should never change after the fact. A migration copies them as they are; a difference between the source and the copy points to a fault in the transfer, not to a correction.
Common errors and how to fix them
Troubleshooting: NONE is a valid value and means no variant applies; it is not the same as a missing attribute, which makes the MR image invalid because the attribute is Type 1.
Check a file in seconds with the free DICOM Validator: it reads the header in your browser, lists the tags with their values, and flags missing or malformed key attributes.
Sequence Variant: frequently asked questions
What is DICOM tag (0018,0021) Sequence Variant?
Variants of the scanning sequence, such as segmented k-space, spoiled or magnetization-prepared.
What VR and VM does Sequence Variant (0018,0021) have?
(0018,0021) SequenceVariant has Value Representation CS (Code String) and Value Multiplicity 1-n, which means one or more values, separated by a backslash (\). A short code. Leading and trailing spaces are not significant. Codes are usually Defined Terms or Enumerated Values listed by the standard. Maximum length: 16 bytes maximum.
Is Sequence Variant (0018,0021) required?
In the MR Image Module (PS3.3 C.8.3.1) it is Type 1: required, and it must have a value.
How is Sequence Variant handled when de-identifying DICOM files?
Sequence Variant is not listed in PS3.15 Table E.1-1, so the Basic Application Level Confidentiality Profile gives no specific action for it.
How do I troubleshoot Sequence Variant (0018,0021)?
NONE is a valid value and means no variant applies; it is not the same as a missing attribute, which makes the MR image invalid because the attribute is Type 1.
Check SequenceVariant in your files
The free DICOM Validator lists every header tag and flags problems. Files stay in your browser.
Open DICOM Validator →Remove patient data before sharing
The free DICOM Anonymizer applies a de-identification profile in your browser, with no upload.
Open DICOM Anonymizer →Sources
Checked against the DICOM standard, release 2026d. Module Types, query keys and de-identification actions come from these tables: