(0008,0030)
Study Time
Keyword: StudyTime
Time the study started, in the format HHMMSS.FFFFFF.
Tag Attributes
- Tag ID
- (0008,0030)
- Compact
- 00080030
- Keyword
- StudyTime
- VR
- TM — Time
- VM
- 1
- Required
- Type 2 — required, may be empty
- PS3.15
- Basic Profile action Z
- Retired
- No
Defined in
What Study Time contains
Combined with Study Date for precise temporal ordering. Critical for studies with multiple exams on the same day.
Group 0008 holds general identifying information: what kind of object this is, when it was made, and by which equipment and people.
Value format: TM (Time), VM 1
(0008,0030) StudyTime has Value Representation TM (Time) and Value Multiplicity 1, which means exactly one value.
TM — Time
A time of day on a 24-hour clock in the format HHMMSS.FFFFFF. Components on the right may be left out, so 14, 1425 and 142530.5 are all valid.
- Length
- 14 bytes maximum
- Characters
- The digits 0–9 and the period.
Example value
142530
Format mistakes to avoid
- Colons (14:25:30) come from the old ACR-NEMA format and are not valid in DICOM.
- A time carries no time zone. If one is needed, Timezone Offset From UTC (0008,0201) holds it.
Where StudyTime appears
| Defined in | Type | Meaning |
|---|---|---|
| General Study ModulePS3.3 C.7.2.1 | 2 | Required, but it may be empty (zero length) when the value is unknown. |
A module lists the attributes an object must or may carry. Many IODs (CT Image, MR Image, Secondary Capture and others) include the same module, and an IOD-specific module can make an attribute stricter than the general one.
In a C-FIND query using the Study Root Query/Retrieve Information Model, it is a STUDY-level key of type R: a Required Key: every Query/Retrieve SCP must support matching on it.
De-identification (PS3.15)
PS3.15 Table E.1-1 gives Study Time the Basic Profile action Z: replace it with an empty value, or with a dummy value consistent with the VR.
With the Retain Longitudinal Temporal Information with Full Dates Option, the action is K: keep it unchanged (a sequence is kept but cleaned item by item).
With the Retain Longitudinal Temporal Information with Modified Dates Option, the action is C: clean it: replace values with similar values that carry no identifying information.
To see how de-identification changes your own files, try the free DICOM Anonymizer, which runs in the browser and uploads nothing.
In archives and migrations
Study-level attributes are what people search on, so a migration should keep them identical. When the same study exists in two archives with different study-level values, users see two studies, or none, depending on which values they search for.
Common errors and how to fix them
Troubleshooting: Seconds and fractions are optional, so 1425, 142530 and 142530.125 are all valid and should be compared as times, not as strings. Use the same clock source (NTP) on all devices so that times from different systems line up.
Check a file in seconds with the free DICOM Validator: it reads the header in your browser, lists the tags with their values, and flags missing or malformed key attributes.
Study Time: frequently asked questions
What is DICOM tag (0008,0030) Study Time?
Time the study started, in the format HHMMSS.FFFFFF.
What VR and VM does Study Time (0008,0030) have?
(0008,0030) StudyTime has Value Representation TM (Time) and Value Multiplicity 1, which means exactly one value. A time of day on a 24-hour clock in the format HHMMSS.FFFFFF. Components on the right may be left out, so 14, 1425 and 142530.5 are all valid. Maximum length: 14 bytes maximum.
Is Study Time (0008,0030) required?
In the General Study Module (PS3.3 C.7.2.1) it is Type 2: required, but it may be empty (zero length) when the value is unknown.
Can you query a PACS by Study Time?
In a C-FIND query using the Study Root Query/Retrieve Information Model, it is a STUDY-level key of type R: a Required Key: every Query/Retrieve SCP must support matching on it.
How is Study Time handled when de-identifying DICOM files?
PS3.15 Table E.1-1 gives Study Time the Basic Profile action Z: replace it with an empty value, or with a dummy value consistent with the VR. With the Retain Longitudinal Temporal Information with Full Dates Option, the action is K: keep it unchanged (a sequence is kept but cleaned item by item). With the Retain Longitudinal Temporal Information with Modified Dates Option, the action is C: clean it: replace values with similar values that carry no identifying information.
How do I troubleshoot Study Time (0008,0030)?
Seconds and fractions are optional, so 1425, 142530 and 142530.125 are all valid and should be compared as times, not as strings. Use the same clock source (NTP) on all devices so that times from different systems line up.
Check StudyTime in your files
The free DICOM Validator lists every header tag and flags problems. Files stay in your browser.
Open DICOM Validator →Remove patient data before sharing
The free DICOM Anonymizer applies a de-identification profile in your browser, with no upload.
Open DICOM Anonymizer →Sources
Checked against the DICOM standard, release 2026d. Module Types, query keys and de-identification actions come from these tables: