(0020,0037)
Image Orientation (Patient)
Keyword: ImageOrientationPatient
Direction cosines of the first row and first column of the image with respect to the patient. Six values: row_x, row_y, row_z, col_x, col_y, col_z.
Tag Attributes
- Tag ID
- (0020,0037)
- Compact
- 00200037
- Keyword
- ImageOrientationPatient
- VR
- DS — Decimal String
- VM
- 6
- Required
- Type 1 — required, with a value
- PS3.15
- Not listed in PS3.15 Table E.1-1
- Retired
- No
Defined in
What Image Orientation (Patient) contains
Image Orientation (Patient) holds six direction cosines: the first three give the direction of the rows (moving right along a row of the image), the last three the direction of the columns (moving down), both in the patient coordinate system where x points to the patient's left, y to the back and z to the head.
Viewers derive the orientation labels at the image edges from these values, and their cross product gives the slice normal used to sort slices and compute spacing.
The two vectors are expected to be of unit length and perpendicular to each other. Values written with too few decimals can break that slightly, which strict software reports.
Defines the image plane in patient 3D coordinates. Combined with Image Position (Patient), fully specifies the image plane. Critical for PACS orientation labels and 3D rendering.
Group 0020 holds the identifiers and numbers that tie instances into series and studies, and the geometry of the patient coordinate system.
Value format: DS (Decimal String), VM 6
(0020,0037) ImageOrientationPatient has Value Representation DS (Decimal String) and Value Multiplicity 6, which means exactly six values, separated by backslashes (\).
DS — Decimal String
A fixed-point or floating-point number written as text, for example 0.5, -1024 or 1.2E-3. Leading and trailing spaces are allowed.
- Length
- 16 bytes maximum
- Characters
- The digits 0–9, +, -, E, e, the period and the space.
Example value
1\0\0\0\1\0
Format mistakes to avoid
- Writing a number at full floating-point precision can exceed 16 characters, which is invalid.
- A comma as decimal separator (0,5) is invalid; DICOM always uses the period.
Common values
| Value | Meaning |
|---|---|
| 1\0\0\0\1\0 | Axial: rows run toward patient left, columns toward posterior |
| 1\0\0\0\0\-1 | Coronal: rows toward left, columns toward feet |
| 0\1\0\0\0\-1 | Sagittal: rows toward posterior, columns toward feet |
Where ImageOrientationPatient appears
| Defined in | Type | Meaning |
|---|---|---|
| Image Plane ModulePS3.3 C.7.6.2 | 1 | Required, and it must have a value. |
A module lists the attributes an object must or may carry. Many IODs (CT Image, MR Image, Secondary Capture and others) include the same module, and an IOD-specific module can make an attribute stricter than the general one.
De-identification (PS3.15)
Image Orientation (Patient) is not listed in PS3.15 Table E.1-1, so the Basic Application Level Confidentiality Profile gives no specific action for it.
To see how de-identification changes your own files, try the free DICOM Anonymizer, which runs in the browser and uploads nothing.
In archives and migrations
Image geometry attributes are copied unchanged in a migration. Viewers rebuild 3D volumes, MPR and measurements from them, so a changed value changes the clinical picture.
Common errors and how to fix them
Left and right are swapped on every image of a scan
Usually the wrong patient position was selected at the console, so the scanner computed wrong orientations. The images have to be corrected or re-acquired; editing labels does not fix the geometry.
A reconstruction from gantry-tilted CT looks sheared
With gantry tilt, the line through the slice positions is not parallel to the slice normal. Software has to use the actual positions instead of assuming the slices are stacked along the normal.
Troubleshooting: Wrong orientation labels (e.g., Anterior/Posterior swapped) in PACS are caused by incorrect Image Orientation (Patient) values. Verify that direction cosines match the acquisition plane.
Check a file in seconds with the free DICOM Validator: it reads the header in your browser, lists the tags with their values, and flags missing or malformed key attributes.
Image Orientation (Patient): frequently asked questions
What is DICOM tag (0020,0037) Image Orientation (Patient)?
Direction cosines of the first row and first column of the image with respect to the patient. Six values: row_x, row_y, row_z, col_x, col_y, col_z. Image Orientation (Patient) holds six direction cosines: the first three give the direction of the rows (moving right along a row of the image), the last three the direction of the columns (moving down), both in the patient coordinate system where x points to the patient's left, y to the back and z to the head.
What VR and VM does Image Orientation (Patient) (0020,0037) have?
(0020,0037) ImageOrientationPatient has Value Representation DS (Decimal String) and Value Multiplicity 6, which means exactly six values, separated by backslashes (\). A fixed-point or floating-point number written as text, for example 0.5, -1024 or 1.2E-3. Leading and trailing spaces are allowed. Maximum length: 16 bytes maximum.
Is Image Orientation (Patient) (0020,0037) required?
In the Image Plane Module (PS3.3 C.7.6.2) it is Type 1: required, and it must have a value.
How is Image Orientation (Patient) handled when de-identifying DICOM files?
Image Orientation (Patient) is not listed in PS3.15 Table E.1-1, so the Basic Application Level Confidentiality Profile gives no specific action for it.
How do I troubleshoot Image Orientation (Patient) (0020,0037)?
Wrong orientation labels (e.g., Anterior/Posterior swapped) in PACS are caused by incorrect Image Orientation (Patient) values. Verify that direction cosines match the acquisition plane.
Check ImageOrientationPatient in your files
The free DICOM Validator lists every header tag and flags problems. Files stay in your browser.
Open DICOM Validator →Remove patient data before sharing
The free DICOM Anonymizer applies a de-identification profile in your browser, with no upload.
Open DICOM Anonymizer →Sources
Checked against the DICOM standard, release 2026d. Module Types, query keys and de-identification actions come from these tables: