MajwareMAJWARE
Dose

(0018,9346)

CTDI Phantom Type Code Sequence

Keyword: CTDIPhantomTypeCodeSequence

The type of phantom used for the CTDI measurement, as one coded Item from CID 4053 (CTDI Phantom Device).

Tag Attributes

Tag ID
(0018,9346)
Compact
00189346
Keyword
CTDIPhantomTypeCodeSequence
VR
SQ — Sequence of Items
VM
1
Required
Type 3 — optional
PS3.15
Not listed in PS3.15 Table E.1-1
Retired
No

Defined in

CT Image Module

What CTDI Phantom Type Code Sequence contains

CTDIvol values for the 16 cm head phantom and the 32 cm body phantom differ substantially for the same settings (roughly by a factor of two), so a CTDIvol value can only be compared with others when the phantom is known. Only one Item is allowed.

Group 0018 holds acquisition information: how the data were acquired and with which settings.

Value format: SQ (Sequence of Items), VM 1

(0018,9346) CTDIPhantomTypeCodeSequence has Value Representation SQ (Sequence of Items) and Value Multiplicity 1, which means one sequence, which can hold any number of Items.

SQ — Sequence of Items

A container holding zero or more Items, each of which is a small nested data set with its own attributes.

Length
Not applicable to the value; a sequence is encoded with an explicit length or with delimiters
Characters
Nested data elements, encoded with the same transfer syntax as the rest of the data set.

What an Item contains

Item 1:
  (0008,0100) Code Value                113691
  (0008,0102) Coding Scheme Designator  DCM
  (0008,0104) Code Meaning              IEC 320mm Body CT Dosimetry Phantom

Format mistakes to avoid

  • Software that reads only top-level attributes misses the values inside the Items.
  • De-identification has to go into every Item, because identifying values can sit inside nested sequences.

Common values

ValueMeaning
113690 (DCM)IEC 160mm Head CT Dosimetry Phantom
113691 (DCM)IEC 320mm Body CT Dosimetry Phantom
130541 (DCM)100 mm Pediatric Head CT Dosimetry Phantom

Where CTDIPhantomTypeCodeSequence appears

Defined inTypeMeaning
CT Image ModulePS3.3 C.8.2.13Optional.

A module lists the attributes an object must or may carry. Many IODs (CT Image, MR Image, Secondary Capture and others) include the same module, and an IOD-specific module can make an attribute stricter than the general one.

De-identification (PS3.15)

CTDI Phantom Type Code Sequence is not listed in PS3.15 Table E.1-1, so the Basic Application Level Confidentiality Profile gives no specific action for it.

To see how de-identification changes your own files, try the free DICOM Anonymizer, which runs in the browser and uploads nothing.

In archives and migrations

Dose values are copied unchanged with the images. For dose registries and audits, the Radiation Dose Structured Report, a separate object, is the usual source, so check that those objects are migrated as well.

Common errors and how to fix them

Troubleshooting: When comparing CTDIvol across scanners or protocols, for example paediatric heads against adult bodies, group the values by phantom first. An absent sequence means the phantom is not stated in the image, not that a default applies.

Check a file in seconds with the free DICOM Validator: it reads the header in your browser, lists the tags with their values, and flags missing or malformed key attributes.

CTDI Phantom Type Code Sequence: frequently asked questions

What is DICOM tag (0018,9346) CTDI Phantom Type Code Sequence?

The type of phantom used for the CTDI measurement, as one coded Item from CID 4053 (CTDI Phantom Device).

What VR and VM does CTDI Phantom Type Code Sequence (0018,9346) have?

(0018,9346) CTDIPhantomTypeCodeSequence has Value Representation SQ (Sequence of Items) and Value Multiplicity 1, which means one sequence, which can hold any number of Items. A container holding zero or more Items, each of which is a small nested data set with its own attributes. Maximum length: Not applicable to the value; a sequence is encoded with an explicit length or with delimiters.

Is CTDI Phantom Type Code Sequence (0018,9346) required?

In the CT Image Module (PS3.3 C.8.2.1) it is Type 3: optional.

How is CTDI Phantom Type Code Sequence handled when de-identifying DICOM files?

CTDI Phantom Type Code Sequence is not listed in PS3.15 Table E.1-1, so the Basic Application Level Confidentiality Profile gives no specific action for it.

How do I troubleshoot CTDI Phantom Type Code Sequence (0018,9346)?

When comparing CTDIvol across scanners or protocols, for example paediatric heads against adult bodies, group the values by phantom first. An absent sequence means the phantom is not stated in the image, not that a default applies.

Check CTDIPhantomTypeCodeSequence in your files

The free DICOM Validator lists every header tag and flags problems. Files stay in your browser.

Open DICOM Validator →

Remove patient data before sharing

The free DICOM Anonymizer applies a de-identification profile in your browser, with no upload.

Open DICOM Anonymizer →